Calibration and Maintenance Strategies for Granite Surface Plates

Keywords: Granite surface plate calibration, autocollimator, laser interferometer, uncertainty evaluation
Proper calibration and maintenance are critical to preserving plate accuracy:

  1. Calibration Methods:
    • Autocollimators: Provide angular measurements with 0.05″ resolution.
    • Laser Systems: Enable dynamic flatness profiling via interferometric analysis.
    • CMMs: Use tactile probing for point-by-point verification.
      A study comparing these methods found autocollimators fastest (5 min/plate) but CMMs most accurate (<0.3μm uncertainty).
  2. Maintenance Protocols:
    • Cleaning: Use pH-neutral detergents to avoid chemical etching.
    • Storage: Maintain ambient temperatures (20±2°C) and humidity (45-55% RH).
    • Recalibration: Biannual checks are recommended; plates in high-use environments may require quarterly validation.

Uncertainty budgets, calculated using GUM principles, should include contributions from measurement equipment, environmental factors, and operator variability.